BlogsAsk The Experts ForumWhich Frequency Band Is Most Suitable if Passive Inductive Coupling Within 3 Meters Is Required?

Which Frequency Band Is Most Suitable if Passive Inductive Coupling Within 3 Meters Is Required?

Posted By RFID Journal, 01.21.2011
To reduce the amount of RF interference within a medical environment, an RFID system should have as low a power density as possible. High-frequency and above are feasible—but which requires the least power to communicate between tags and readers?

—Name withheld

———
I reached out to Victor Vega at NXP Semiconductors, a provider of both passive high-frequency (HF) and ultrahigh-frequency (UHF) technology. According to Victor, inductive coupling with passive low-frequency (LF), HF and UHF can not be accomplished at 3 meters (10 feet). "It is possible with active components," he says. "I designed an LF active tag many years ago that had a read range of about 3 meters, but it required a battery."

Victor notes that some interference tests conducted on medical equipment indicated passive RFID had an impact on older EKG equipment—not that the equipment didn't work, but rather there was noise in the EKG signals. "It turned out that the signal cables were unshielded," he explains, "and they resolved that problem by simply using more updated shielding. To further reduce any potential interference, they simply specified a 'safe' distance (or reader power level), at which they tested and found no interference."

I would add that I have inquired about interference with companies that have used RFID technologies in hospitals—both passive and active—and no one has indicated that they have seen any signs that the technology had an impact on their hospital equipment, though a paper published in the
Journal of the American Medical Association (JAMA) did find that passive RFID readers can interfere with equipment (see Researchers Warn RFID May Disrupt Medical Equipment).

AIM Global has been working on guidelines to ensure the safe operation of medical equipment around RFID systems in a hospital setting. If you have any concerns about this, I would suggest choosing the RFID system that best meets your needs, and then testing it with equipment used in the same area.

—Mark Roberti, Founder and Editor,
RFID Journal

Login and post your comment!

Not a member?

Signup for an account now to access all of the features of RFIDJournal.com!

Next Post
Is There a "Gold Standard" for 13.56 MHz...
Previous Post
How Can an RFID Chip Be Implanted in the...
PREMIUM CONTENT
Case Studies Features Best Practices How-Tos
RFID JOURNAL EVENTS
Live Events Virtual Events Webinars
ASK THE EXPERTS
Simply enter a question for our experts.
TAKE THE POLL
JOIN THE CONVERSATION ON TWITTER
Loading
RFID Journal LIVE! RFID in Health Care LIVE! LatAm LIVE! Brasil LIVE! Europe RFID Connect Virtual Events RFID Journal Awards Webinars Presentations
© Copyright 2002-2016 RFID Journal LLC.
Powered By: Haycco